• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND GMBH S5 CONTROL PANEL MRP 85; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND GMBH S5 CONTROL PANEL MRP 85; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 28-95-85
Device Problems Unstable (1667); Detachment of Device or Device Component (2907)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/30/2020
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 control panel mrp 85.The incident occurred in (b)(6).Through follow-up communication livanova learned that at the pre-test before going in the procedure the knob was acting normally and that it became loose during the case.However, the pump speed was still controllable by the knob and the procedure could be completed without issues.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a s5 control panel mrp 85 speed control knob was wobbling during procedure.After the procedure, it was checked and completely detached from the panel.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 CONTROL PANEL MRP 85
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND GMBH
lindberghstr. 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key9764262
MDR Text Key197121322
Report Number9611109-2020-00143
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K071318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 02/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/27/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number28-95-85
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/31/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/13/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-