Model Number N/A |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
Foreign Body In Patient (2687)
|
Event Date 12/17/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Concomitant medical product(s): 00780502400 neck provisional r 60796174.Customer has indicated that the product will not be returned to zimmer biomet for investigation.The product location is unknown.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001822565 - 2020 - 00693, trial neck.
|
|
Event Description
|
It was reported during a routing trial head reduction, the trial head came off the trial m/l taper and was lost in the wound.The trial head was located, but it was too deep in the tissues for retrieval to be made.The surgeon finally decided to complete the total hip procedure and leave the trial femoral head in place until procedure to remove could be scheduled for the following day.This caused a 90 minute surgical delay.There is no additional information available.
|
|
Manufacturer Narrative
|
Product was not returned for evaluation.Lot identification is necessary for review of device history records, lot identification was not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No additional information reported.
|
|