A steris service technician arrived onsite to inspect the examiner 10 light.While onsite, the technician was informed by facility personnel that they contacted steris to request functional testing of the examiner 10 light in an attempt to rule out the light as a root cause of the reported burn.The technician performed an inspection and found the light to be operating properly.No issue with the function or operation of the light was identified; no repairs were required.Steris understands that the hospital is conducting an investigation of the reported event, including the use of a topical treatment that was administered to the patient prior to the start of the procedure as a potential root cause.The hospital has not provided steris further information regarding the topical treatment applied.No additional issues have been reported.
|