• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH CONFIDENCE KIT, NO NEEDLES; POLYMETHYLMETHACRYLATE BONE CEMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH CONFIDENCE KIT, NO NEEDLES; POLYMETHYLMETHACRYLATE BONE CEMENT Back to Search Results
Model Number 283913000
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 11/27/2019
Event Type  malfunction  
Manufacturer Narrative
A product investigation was conducted.Visual inspection: upon visual inspection, it was observed that only the hydraulic pump was returned from the confidence kit and there was some moisture observed inside the pump.No other issues were identified with the returned device.Functional test: the functional test cannot be performed on the returned device as the hydraulic pump was returned by itself.Dimensional inspection: a dimensional inspection was not performed as the internal components were inaccessible without destruction of the device.Document/specification review: based on the date of manufacture the relevant drawings, reflecting the current and manufactured revision were reviewed yes, the device received was leaking.Hence confirming the allegation.Investigation conclusion: the complaint condition was confirmed for the confidence kit, no needles (p/n: 283913000, lot #: 244229).There is no indication that a design or manufacturing issue has caused the issue and hence the root cause cannot be determined.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history record (dhr) review was conducted: the dhr of product code 283913000, lot 244229, was reviewed and no non-conformances were observed during the manufacturing process.The product was released on june 11th, 2019.Qty.(b)(4).The dhr was electronically reviewed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Device report from synthes reports an event in (b)(6) as follows: it was reported that during an unknown procedure on (b)(6) 2019, there was a pump issue.Surgeon used another same type of product to complete the procedure successfully.There was no patient consequences reported.During manufacturer's investigation of the returned device it was observed that only the hydraulic pump was returned from the confidence kit and there was some moisture observed inside the pump.This is report 1 of 1 for (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONFIDENCE KIT, NO NEEDLES
Type of Device
POLYMETHYLMETHACRYLATE BONE CEMENT
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle 02400
SZ   02400
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key9772178
MDR Text Key203903391
Report Number1526439-2020-00634
Device Sequence Number1
Product Code NDN
UDI-Device Identifier10705034209630
UDI-Public(01)10705034209630
Combination Product (y/n)N
Reporter Country CodePL
PMA/PMN Number
K060300
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial
Report Date 12/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received02/28/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number283913000
Device Catalogue Number283913000
Device Lot Number244229
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/10/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/26/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/11/2019
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
-
-