It was reported that during surgery, when the 40 degree chondral pick was used, the device was brittle and the tip of it bent easily 90 degree.It is unknown whether there was a delay in the case.The procedure was completed with a competitor device.No patient injuries reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
H10 h3, h6: the reported 40 degree chondral pick, used in treatment, has been returned for evaluation.Visual assessment of the device confirmed the reported complaint.The distal tip of the pick has been bent to 90 degrees.The material condition was tested and confirmed to be within print specification.The condition of the device indicates it was subjected to excessive force during use.Per the devices instruction for use ¿excessive force should not be applied to the instrument when manipulating soft tissue, bone, or hard objects.Misuse of these instruments may result in bent distal tips¿.A review of the manufacturing and complaint records was performed for the reported lot, there were no indications that would suggest that the device did not meet product specifications upon release into distribution.
|