It was reported that during use in a patient, while inserting the swan-ganz catheter, resistance was encountered on insertion past ~30cm.The balloon was deflated and the pacing catheter was pulled back.The physician attempted again to float the pacing catheter but was the catheter was unable to pass.Upon removing pacing catheter from the introducer, the catheter was noted to be kinked.This resulted in delay of pacing secondary to the physician reestablishing a sterile field and obtaining another pacing catheter for insertion.The subsequent pacer floated with no issue.There was no allegation of patient injury.The device was available for evaluation.
|
One d97120f5 swan-ganz catheter with an attached non-ew contamination shield and 1.3cc syringe were returned for examination.The reported event of "catheter noted to be kinked" was confirmed.The non-ew contamination shield was received attached to the catheter.The contamination shield was removed, and an indentation was observed in the catheter body at approximately 28 cm from the catheter tip, where distal connector of the returned contamination shield was located.The returned contamination shield was able to be attached to the returned catheter without any difficulties.The outer diameter of the catheter body was measured and found to be within specification.Continuity testing was performed on the distal and proximal circuits and there were no open, intermittent, or short conditions observed.The balloon inflated clear and concentric with 1.3 cc air and the balloon remained inflated for 5 minutes without leakage.A review of the manufacturing records indicated that the product met specifications upon release.An engineering evaluation has been initiated to assess for any manufacturing-related processes which could be correlated to the complaint.Swan-ganz pacing thermodilution (td) catheters serve as diagnostic and therapeutic tools in the management of critically ill patients.There are multiple failure modes that may require the exchange of a pacing catheter.Since proper functioning of the pacing catheter depends on the electrical continuity of its electrodes and internal wires, care should be exercised when handling the catheter.Stretching, kinking, or forceful wiping of the catheter may result in damage.After stable pacing has been confirmed, the proximal end of the catheter should be secured to the insertion site to prevent undue movement that could result in tip dislodgment and loss of capture, or catheter migration.Care should be taken not to kink the catheter body when securing it.In this complaint, it could not be determined if procedural factors or device handling may have contributed to the reported event.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|