Following the end of a cryoablation procedure, ablation with radiofrequency was attempted in the cavotricuspid isthmus.The equipment was connected according to indication, however when the generator was switched on, an error message displayed ¿generator malfunction.Contact sjm technical support ¿.The equipment was turned off and reconnected, but the issue remained.The ablation was not continued and there were no patient consequences.
|
One ampere¿ rf ablation generator was received for investigation.When power was applied to the generator, the screen illuminated red while displaying a generator malfunction error, confirming the reported event.The rf/mn (radiofrequency/matching network) assembly was temporarily replaced with a known good assembly and upon power cycle to the generator, a normal boot up sequence was observed.All connector ports were then tested for functionality, numerous catheter codes were manually applied, various impedance & temperature values were also applied for investigation upon which all values were accurately tracked on the generator¿s display screen, no abnormal impedance values were observed.The rf energy output was measured during ablation testing and no further anomalies were identified throughout investigation.Based on the information provided to abbott and the investigation performed, the root cause of the field reported event was successfully isolated to abnormal functionality of the rf/mn assembly.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.
|