• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAQUET CARDIOPULMONARY GMBH HEART LUNG MACHINE; PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE Back to Search Results
Model Number 701043292 - ROTAFLOW ENGLISH/US US-PLUG
Device Problem Device Displays Incorrect Message (2591)
Patient Problem No Patient Involvement (2645)
Event Date 02/24/2020
Event Type  malfunction  
Manufacturer Narrative
The reported failure "head error " occurred in the service center during maintenance.According to the service order (b)(4) dated on 2020-02-25 rotaflow console has a bad control board.Replaced control board ((b)(4)) and tested per service manual.All test passed.Most possible root cause could be determined as: the head error is caused by the hot plug.When device is in operation and the power plug is plugged in or out.And this leads to a damage at the control board of the rotaflow.The head error follows by the sig error.This is when the ultrasonic crème is applied to the flow bubble sensor.Then the centrifugal pump is causing backflow and this leads to the head error.The head error is also caused by shaking the drive.This is when the motor (which is controlled by the optical tacho) is not blocked when adjusting to 0 then it could lead to error when slight shaking.The motor could then slightly rotate.The head error can be caused by connection issues between the console (rfc) and the drive (rfd).This is when the cable connection is disturbed by defective pins.Furthermore, the instructions for use of the rotaflow system, see rotaflow system user manuel, mcv-ga-10000703-de-11 contain detailed descriptions to prevent an ¿error head¿.The reported failure "head error" occured during service and could be confirmed.The device was directly involved in the incident.The occurrence rate regarding the above complaint is below the acceptance rate.Thus, no remedial action required.The occurrence rate related to the reported issue is currently being monitored as part of maquet cardiopulmonary¿s trending program and additional investigations or corrections will be implemented in case of adverse trending.
 
Event Description
It was reported from the repair center us that head error on rotaflow console discovered by (b)(6) in the (b)(4) center.No patient involvement.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEART LUNG MACHINE
Type of Device
PUMP, BLOOD, CARDIOPULMONARY BYPASS, NON-ROLLER TYPE
Manufacturer (Section D)
MAQUET CARDIOPULMONARY GMBH
neue rottenburger strasse 37
hechingen
Manufacturer (Section G)
NURSEL BOELENS
maquet cardiopulmonary ag
kehler strasse 31
rastatt 76437
GM   76437
Manufacturer Contact
maquet cardiopulmonary ag
kehler strasse 31
rastatt 76437
4972229321
MDR Report Key9782289
MDR Text Key190542118
Report Number8010762-2020-00090
Device Sequence Number1
Product Code KFM
Combination Product (y/n)N
PMA/PMN Number
K991864
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,foreign,health profe
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/03/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number701043292 - ROTAFLOW ENGLISH/US US-PLUG
Device Catalogue Number701043292
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/28/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/29/2012
Is the Device Single Use? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-