• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE CORPORATION BURETROL SOLUTION SETS; SET, BLOOD TRANSFUSION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE CORPORATION BURETROL SOLUTION SETS; SET, BLOOD TRANSFUSION Back to Search Results
Catalog Number EMC9601N
Device Problem Device Contamination with Chemical or Other Material (2944)
Patient Problem No Patient Involvement (2645)
Event Date 02/07/2020
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported particulate matter was observed in a buretrol blood administration set.The particulate appeared to be ¿piece of paper¿ noted "within the buretrol chamber".The issue was noted during priming of the line.Was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information was added: the device was received for evaluation.Visual inspection revealed particulate matter within the burette.The reported condition was verified.The cause of the condition was determined to be a supplier manufacturing issue.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BURETROL SOLUTION SETS
Type of Device
SET, BLOOD TRANSFUSION
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - TUNISIA
route de chebbaou
2021oued e
tunis
TS  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9793020
MDR Text Key182162788
Report Number1416980-2020-01195
Device Sequence Number1
Product Code BRZ
Combination Product (y/n)N
Reporter Country CodeEI
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 04/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/05/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue NumberEMC9601N
Device Lot Number19G06T037
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/31/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/20/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-