Model Number C408646
Device Problem Gas/Air Leak (2946)
Patient Problem Air Embolism (1697)
Event Date 02/06/2020
Event Type  Injury  
Manufacturer Narrative
One dragonfly optis imaging catheter was received for evaluation.A hoop and the syringe were also returned.The results of the investigation confirmed that the returned catheter was able to be purged as normal, after the dried contrast was removed via water bath, with no air bubbles, leaks, or anomalies noted.Dimensional inspection revealed the purge hole diameter also met specifications.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the returned condition of the device and the information received, the cause of the reported event remains unknown.
 
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
When the catheter was purged, air bubbles were noted in the side port.The bubbles were attempted to be aspirated and elevations were noticed in the ecg.The device was removed and the procedure was completed without oct.Oxygen was administered and the elevations disappeared after a few minutes.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRAGONFLY OPTIS KIT BOX IMAGING CATHETER
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
ST. JUDE MEDICAL CATD
4 robbins drive
westford MA 01886
Manufacturer (Section G)
ST. JUDE MEDICAL CATD
4 robbins drive
westford MA 01886
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key9799370
MDR Text Key188862169
Report Number3009600098-2020-00004
Device Sequence Number1
Product Code DQO
UDI-Device Identifier00183739000654
UDI-Public00183739000654
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K141453
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 04/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date09/22/2021
Device Model NumberC408646
Device Catalogue NumberC408646
Device Lot Number7178915
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/10/2020
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received03/18/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/24/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-