• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: AGA MEDICAL CORPORATION AMPLATZER PICCOLO OCCLUDER; OCCLUDER, PATENT DUCTUS, ARTERIOSUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

AGA MEDICAL CORPORATION AMPLATZER PICCOLO OCCLUDER; OCCLUDER, PATENT DUCTUS, ARTERIOSUS Back to Search Results
Model Number 9-PDAP-04-02-L
Device Problems Gradient Increase (1270); Obstruction of Flow (2423); Migration (4003)
Patient Problem Stenosis (2263)
Event Type  Injury  
Manufacturer Narrative
Further information regarding this event has been requested.The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Event Description
On (b)(6) 2019, a 4/2mm amplatzer piccolo occluder was implanted in a premature baby born at (b)(6) weeks gestation and weighed (b)(6) kilograms.The infant had a history of enterocolitis related to prematurity.The device was successfully placed with no evidence of aortic or pulmonary protrusion.Post implant, an aortic obstruction and increase gradient was reported.One month post implant on (b)(6) 2019, it was reported that the device migrated slightly towards the aorta.The physician decided to continue to monitor the infant.The infant remains in the nicu in stable condition.
 
Manufacturer Narrative
An event of migration of the 04-02 piccolo occluder into the aorta was reported.The results of the investigation are inconclusive since the device was not returned for analysis.A review of the measurements as reported from the field was performed.Per the sizing table in the instructions for use, arten600042307 version a, the correct size piccolo occluder for the measurements provided was 04-02, the size of the implanted device.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Manufacturer Narrative
Correction: d2.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER PICCOLO OCCLUDER
Type of Device
OCCLUDER, PATENT DUCTUS, ARTERIOSUS
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
MDR Report Key9802234
MDR Text Key183949205
Report Number2135147-2020-00107
Device Sequence Number1
Product Code MAE
Combination Product (y/n)N
PMA/PMN Number
P020024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 12/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/06/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number9-PDAP-04-02-L
Device Catalogue Number9-PDAP-04-02-L
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Date Manufacturer Received11/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Weight1
-
-