The ventilator failure described by the user could be reconstructed by means of the logfile analysis.No indications for a device malfunction were found.The autonomous safety shutdown of the ventilator was triggered by a too fast and too high pressure increase, possibly due to the patient breathing or coughing against the ventilator.The device behaved as specified with an autonomous shutdown while changing mode to man/spont (safety mode) accompanied by an audible and visible "ventilator fail" alarm.At the time in question, the volume af mode was used; in general, it is recommended to switch on the synchronization for spontaneously breathing patients (e.G.Volume af mode with activated trigger) or use synchronized ventilation mode (e.G.Pressure support).The device behaved as specified with an autonomous shutdown while changing mode to man/spont (safety mode) accompanied by an audible and visible "ventilator fail" alarm.The number of similar cases, related to the same root cause, is within the expected range of the respective risk assessment and thus accepted.
|