• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TIBIAL ARTICULAR SURFACE PROVISIONAL RIGHT SIZE GH; PROSTHESIS, KNEE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TIBIAL ARTICULAR SURFACE PROVISIONAL RIGHT SIZE GH; PROSTHESIS, KNEE Back to Search Results
Model Number N/A
Device Problem Fracture (1260)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that the fractured instrument was found in a worn instrument bin.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Verified item lot combination and reviewed manufacturing date as returned item # 42527000707 lot # 63769473 exhibits signs of repeated use and the post/tab has partially fractured on the anterior side.Not all pieces were returned.The complaint is confirmed.The dhr was reviewed and no discrepancies relevant to the reported event were found.Evaluation of the returned device identified the fracture was consistent with the tasp fractures analyzed in a zrm.The zrm identified that the common failure modes for the tasp devices include either bending overload or low cycle fatigue culminating in bending overload as evident by the presence of hackle marks, river lines and striations features on the fracture surface.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TIBIAL ARTICULAR SURFACE PROVISIONAL RIGHT SIZE GH
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9806002
MDR Text Key182613662
Report Number0001822565-2020-00872
Device Sequence Number1
Product Code MBH
Combination Product (y/n)N
PMA/PMN Number
K172524
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Type of Report Initial,Followup
Report Date 03/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/09/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42527000707
Device Lot Number63769473
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer02/18/2020
Was the Report Sent to FDA? No
Date Manufacturer Received03/10/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-