It was reported that within minutes of beginning support the perfusionist reported that the flow dropped from 4l to 2l and eventually 0l while the patient was on the or table.The pump and oxygenator were exchanged.Device history records were reviewed for the returned pump and oxygenator.Both were reviewed and no conformities or abnormalities were found.The pump and oxygenator were received by the manufacturer for product analysis.When the pump was returned a thrombus was evident on the impeller vanes and the current draw values were higher than specified, potentially due to the thrombus observed.It cannot be determined whether the thrombus was present during use in the field, or whether it was a result of return shipping conditions.During internal visual inspection of the pump lower housing and journal bores, particulate was found which may indicate improper priming of the pump during use.The drop in flow was likely due to an increased resistance of the oxygenator due to thrombus formation.No further relevant information has been received to date.
|