Catalog Number 5C4482
Device Problems Delivered as Unsterile Product (1421); Unsealed Device Packaging (1444)
Patient Problem No Patient Involvement (2645)
Event Date 12/26/2019
Event Type  malfunction  
Manufacturer Narrative
Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the sterile package of a minicap transfer set was damaged.This was noticed before the nurse opened the package and prior to use for peritoneal dialysis therapy.There was no patient involvement.No additional information is available.
 
Manufacturer Narrative
Additional information: the device was received for evaluation.A visual inspection was performed with the naked eye noted an opened sterilized pouch.There was evidence of transfer on the clear side of the pouch indicating it was sealed during assembly.The reported condition was verified.The cause of the reported condition could not be determined.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MINICAP TRANSFER SET
Type of Device
SET, ADMINISTRATION, FOR PERITONEAL DIALYSIS, DISPOSABLE
Manufacturer (Section D)
BAXTER HEALTHCARE CORPORATION
deerfield IL
Manufacturer (Section G)
BAXTER HEALTHCARE - MOUNTAIN HOME
1900 n highway 201
mountain home AR 72653
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key9820712
MDR Text Key183086165
Report Number1416980-2020-01349
Device Sequence Number1
Product Code KDJ
UDI-Device Identifier00085412007731
UDI-Public(01)00085412007731
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K152675
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Nurse
Type of Report Initial,Followup
Report Date 04/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/11/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/02/2024
Device Catalogue Number5C4482
Device Lot NumberH19E02038
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/09/2020
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/15/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/02/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown