• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. RETROGRADE DRL 8MM; BIT, SURGICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. RETROGRADE DRL 8MM; BIT, SURGICAL Back to Search Results
Catalog Number 72204042
Device Problem Mechanical Jam (2983)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/17/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(6).
 
Event Description
It was reported that during surgery, when using the acufex trunav retrograde drill and trying to pull the cutting blade back into the shaft, it did not move.So, surgeon had to drill all the way through the tibia and had to use a medium xtendobutton round instead of the surgeon's normal technique (all inside drilling).It is unknown if there was a delay in the procedure.Procedure was changed, but it was completed with the same device.No patient injuries were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10 h3,h6: the reported 8mm trunav retrograde drill, used on treatment, has not been returned for evaluation.Without the reported product a visual evaluation cannot be performed and the customers complaint cannot be confirmed.From the information provided, the cutting blade would not retract.An exact root cause cannot be determined with confidence; however, factors that could have contributed to the reported event include: not retracting the guidewire prior to deployment of the cutting blade.Not having sufficient clearance to allow the cutting blade to deploy properly.The instruction for use outlines precautionary statements and instructions in regards to the use of the device to avoid damage or non-functionality.A review of the manufacturing and complaint records was performed, there were no indications that would suggest that the device did not meet product specifications upon release into distribution.Further investigation is not warranted at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RETROGRADE DRL 8MM
Type of Device
BIT, SURGICAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes boulevard
mansfield MA 02048
MDR Report Key9823487
MDR Text Key183112847
Report Number1219602-2020-00511
Device Sequence Number1
Product Code GFG
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Type of Report Initial,Followup
Report Date 04/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/12/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/29/2024
Device Catalogue Number72204042
Device Lot Number4821694
Was Device Available for Evaluation? No
Date Manufacturer Received04/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-