DEPUY ORTHOPAEDICS INC US QUICKSET 1PC FLEX DRILL BIT 45; DRILLS (BITS, SLEEVES, GUIDES) : DRILLS Back to Search Results
Model Number 2274-45-500
Device Problems Break (1069); Difficult to Remove (1528); Device-Device Incompatibility (2919); Physical Resistance/Sticking (4012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 01/01/2020
Event Type  malfunction  
Manufacturer Narrative
Product complaint # (b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that this complaint is for an instrument 227454500, a drill guide for 6.5 mm screw.Drill bit got stuck in guide.Did not affect surgical outcome, there were several other guides and drill bits sterile and available.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned instrument confirmed the complaint.Drill bit is jammed inside the drill guide.Visual analysis also found that the majority of the fluted tip broke off.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : examination of the returned instrument confirmed the complaint.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary examination of the returned instrument confirmed the complaint.Depuy-synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot null device history batch null device history review null if information is obtained that was not available for the initial medwatch, a follow-up medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSET 1PC FLEX DRILL BIT 45
Type of Device
DRILLS (BITS, SLEEVES, GUIDES) : DRILLS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key9824051
MDR Text Key183137855
Report Number1818910-2020-07601
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10603295109211
UDI-Public10603295109211
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup,Followup
Report Date 02/25/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/12/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2274-45-500
Device Catalogue Number227445500
Device Lot NumberPG250685
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/02/2020
Date Manufacturer Received04/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age64 YR
-
-