SMITH & NEPHEW, INC. MI Z HANDLE ACET REAMER; PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL Back to Search Results
Model Number 71364073
Device Problem Insufficient Information (3190)
Patient Problem Injury (2348)
Event Date 02/20/2020
Event Type  malfunction  
Event Description
It was reported that during a thr procedure the mi z handle acet reamer snapped at the base of the shaft upon reaming.Nothing fell into patient.There were no patient harm and procedural delays reported because there was a backup device available.
 
Manufacturer Narrative
The device, used in treatment was not returned for evaluation, reporting event could not be confirmed.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the device history records for the listed batch did not reveal any deviation from the standard manufacturing processes.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Without the actual product involved and/or device information, our investigation cannot proceed.If the device or new information is received in the future, this complaint can be re-opened.No further actions are being taken at this time.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MI Z HANDLE ACET REAMER
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks road
memphis TN 38116
MDR Report Key9824052
MDR Text Key183130725
Report Number1020279-2020-00841
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010507570
UDI-Public03596010507570
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 06/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/12/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71364073
Device Catalogue Number71364073
Device Lot Number19KM15749
Date Manufacturer Received02/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-