One single dpt kit was returned for examination.The reported event of "foreign object was found inside the fluid path" was not confirmed.A visual examination of the kit both before and after decontamination did not find any presence of a foreign substance inside of the kit.Liquid inside of the kit was also flushed out to filter paper for 5 minutes for further examination.There was no visible foreign substance on the filter paper.However, leakage was detected from the dpt flush device housing.Leakage occurred from one crack, 0.20" approximately in length, below the weld area of the flush device housing.No other visible damage or leakage was observed from the dpt kit.A review of the manufacturing records indicated that the product met specifications upon release.Please note that leakage occurring below the weld area through a crack would not be a reportable event.Leakage implies a small amount of fluid loss that is unlikely to result in an injury.This generally results in the need to exchange the device, which can be done with a minor interruption in therapy.It is common clinical practice to inspect all products before usage.Additionally, these products are used by highly trained clinicians, experienced in identifying and mitigating any hazards that arise.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised to consider the potential benefits in relation to the possible complications.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.
|