EDWARDS LIFESCIENCES EDWARDS INTUITY ELITE VALVE SYSTEM; TISSUE, HEART-VALVE Back to Search Results
Model Number 8300AB
Device Problems Calcified (1077); Mechanical Problem (1384); Structural Problem (2506); Material Deformation (2976)
Patient Problems Aortic Regurgitation (1716); Aortic Valve Stenosis (1717); Regurgitation (2259); Stenosis (2263)
Event Date 12/16/2019
Event Type  Injury  
Manufacturer Narrative
Calcification is a well-recognized failure mode of bioprosthetic valves.Many factors contribute to the onset and propagation of calcification.These include patient factors (age, disease state, pharmacological intervention, etc.), and mechanical stress related to the valve's hemodynamic performance.Though numerous studies have been conducted on preventive calcification strategies in bioprosthetic heart valves, the causes of calcification are not fully understood and there are still no mechanisms or medical therapies which fully prevent bioprostheses from calcifying.Although patient factors are believed to play a crucial role in the development in bioprosthetic tissue calcification, the underline mechanism is still not fully understood.The device was not returned for evaluation, as it remains implanted.The root cause of this event cannot be conclusively determined.However, it is likely that patient related factors and the progression of the underlying valvular disease pathology contributed to the event.The device history record (dhr) was reviewed and showed that this device met all manufacturing and sterilization specifications for product release prior to distribution.No issues were identified that would have impacted this event.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.Udi number: (b)(4).
 
Event Description
It was reported that a 23mm aortic valve was disabled via a valve-in-valve procedure after an implant duration of 2 years due to degeneration, severe stenosis, regurgitation, leaflet thickening, leaflet motion restricted, and calcification.A non-edwards transcatheter valve was implanted.Patient was discharged on pod #1.
 
Manufacturer Narrative
Correction: the g4 date on the initial mdr for the initial aware date should be (b)(6) 2020.
 
Manufacturer Narrative
Corrected data: f10, h6.Reference capa: 20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS INTUITY ELITE VALVE SYSTEM
Type of Device
TISSUE, HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
MDR Report Key9831993
MDR Text Key190957362
Report Number2015691-2020-10961
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
PMA/PMN Number
P150036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,study
Type of Report Initial,Followup,Followup
Report Date 01/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/13/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/26/2021
Device Model Number8300AB
Device Lot NumberR-17H4735
Was Device Available for Evaluation? No
Date Manufacturer Received07/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age71 YR
-
-