BOSTON SCIENTIFIC CORPORATION CRE FIXED WIRE; DILATOR, ESOPHAGEAL Back to Search Results
Model Number M00558370
Device Problems Burst Container or Vessel (1074); Material Deformation (2976)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 12/01/2019
Event Type  malfunction  
Manufacturer Narrative
Date of event was approximated to (b)(6) 2019 as no event date was reported.(b)(4).Investigation results: a visual examination of the returned complaint device found that the balloon burst.It was also noticed that the catheter was kinked.Functional evaluation could not be performed due to the condition of the returned device.Dimensional examination of the catheter was performed and was measured in three sections; distal, medium and proximal.The three sections were within specification.Based on the available information, it is possible that factors encountered during the procedure, the technique used by the physician during the procedure, the amount of strength applied by the customer during the movement of the balloon, and/or the interaction between the scope and the balloon could have resulted in the damage found on the balloon.Therefore, the most probable root cause is adverse event related to procedure.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
 
Event Description
It was reported to boston scientific corporation that a cre fixed wire dilatation balloon was used during a dilation procedure performed in the esophagus on an unknown date.According to the complainant, during the procedure, an attempt to inflate the balloon was made; however, the balloon would not inflate.The procedure was completed with another cre fixed wire dilatation balloon.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be okay.Investigation results revealed that the balloon burst; therefore, this is now an mdr reportable event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CRE FIXED WIRE
Type of Device
DILATOR, ESOPHAGEAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORK LIMITED
cork business technology park
model farm road
cork
EI  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key9835323
MDR Text Key191268761
Report Number3005099803-2020-00832
Device Sequence Number1
Product Code KNQ
UDI-Device Identifier08714729195993
UDI-Public08714729195993
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K971320
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other
Type of Report Initial
Report Date 03/16/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/22/2021
Device Model NumberM00558370
Device Catalogue Number5837
Device Lot Number0024474702
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer01/20/2020
Date Manufacturer Received02/17/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/23/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-