• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. JGRKNT 2.9MM #2 BLUE MB SNGL NGLE; FASTNER, FIXATION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. JGRKNT 2.9MM #2 BLUE MB SNGL NGLE; FASTNER, FIXATION Back to Search Results
Catalog Number 912029
Device Problem Migration or Expulsion of Device (1395)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/14/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).(b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up report will be submitted.
 
Event Description
It was reported that the anchor pulled out during the surgery.Attempts have been made and additional information on the reported event is unavailable.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned product/provided pictures identified the suture is separated from the handle.There is no damage to the handle.Device history record was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JGRKNT 2.9MM #2 BLUE MB SNGL NGLE
Type of Device
FASTNER, FIXATION
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key9866925
MDR Text Key186122437
Report Number0001825034-2020-01219
Device Sequence Number1
Product Code JDR
Combination Product (y/n)N
PMA/PMN Number
K110145
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/05/2023
Device Catalogue Number912029
Device Lot NumberP11959
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/23/2020
Was the Report Sent to FDA? No
Date Manufacturer Received08/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-