Catalog Number 912029 |
Device Problem
Migration or Expulsion of Device (1395)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 02/14/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).(b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up report will be submitted.
|
|
Event Description
|
It was reported that the anchor pulled out during the surgery.Attempts have been made and additional information on the reported event is unavailable.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.Visual examination of the returned product/provided pictures identified the suture is separated from the handle.There is no damage to the handle.Device history record was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|