BOSTON SCIENTIFIC CORPORATION SPEEDBAND SUPERVIEW SUPER 7; LIGATOR, HEMORRHOIDAL Back to Search Results
Model Number M00542251
Device Problem Premature Activation (1484)
Patient Problem Tissue Damage (2104)
Event Date 02/14/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).The device has not been received for analysis.Upon receipt and completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that two speedband superview super 7 devices were used during a trans-oesophageal ligation of oesophageal varices (lvo) procedure performed on (b)(6) 2020.According to the complainant, during the procedure, the handle was rotated in order to deploy an elastic band; however, two elastic bands were able to deploy and happened several times on the same device.The speedband device was removed from the patient, and the second speedband device was used, but the same issue occurred.Reportedly, the deployment failure led to a superficial lesion of the esophageal mucosa which did not require any treatment.The procedure was completed with another speedband superview super 7 device.Additionally, there was no difficulty experienced upon setting up the devices.The patient condition at the conclusion of the procedure was reported to be good.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPEEDBAND SUPERVIEW SUPER 7
Type of Device
LIGATOR, HEMORRHOIDAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key9868254
MDR Text Key191289044
Report Number3005099803-2020-01266
Device Sequence Number1
Product Code FHN
UDI-Device Identifier08714729201960
UDI-Public08714729201960
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
510K EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 03/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/23/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date12/13/2020
Device Model NumberM00542251
Device Catalogue Number4225
Device Lot Number0024930890
Was Device Available for Evaluation? No
Date Manufacturer Received02/28/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/14/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-