MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® ALL-SILICONE FOLEY CATHETER; FOLEY CATHETER (SILICONE)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® ALL-SILICONE FOLEY CATHETER; FOLEY CATHETER (SILICONE) Back to Search Results
Catalog Number 165814
Device Problem Deflation Problem (1149)
Patient Problems No Consequences Or Impact To Patient (2199); Injury (2348); Blood Loss (2597)
Event Type  Injury  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete, a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported that it was difficult to deflate the balloon of the catheter.Reportedly, only 5ml of water was coming out causing trauma and bleeding to the patient.Five patients were admitted to hospital.It is unknown at this time if the patient required medical intervention.
 
Manufacturer Narrative
The reported event was inconclusive as no sample was returned for evaluation.A potential failure mode could be ¿valve stem too long or too short and valve body or stem deformed¿ with a potential root cause of "poor interface/fit between inflation syringe and/ or inflation funnel and incorrect alignment of valve assembly".The lot number is unknown; therefore, the device history record could not be reviewed.The instructions for use were found adequate and state the following: "visually inspect the product for any imperfections or surface deterioration prior to use use luer tip syringe to inflate with stated ml of sterile water.Or ¿ for pre-filled products, remove clip and squeeze reservoir to inflate with stated ml of sterile water.To deflate catheter balloon: gently insert a luer slip tip syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.Allow the pressure within the balloon to force the plunger back and fill the syringe with water.If you notice slow or no deflation, re-seat the syringe gently.Use only gentle aspiration to encourage deflation if needed.Vigorous aspiration may collapse the inflation lumen, preventing balloon deflation.If necessary, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient." section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Event Description
It was reported that it was difficult to deflate the balloon of the catheter.Reportedly, only 5ml of water was coming out causing trauma and bleeding to the patient.Five patients were admitted to hospital.It is unknown at this time if the patient required medical intervention.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® ALL-SILICONE FOLEY CATHETER
Type of Device
FOLEY CATHETER (SILICONE)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key9871423
MDR Text Key187086480
Report Number1018233-2020-02045
Device Sequence Number1
Product Code GBM
Combination Product (y/n)N
PMA/PMN Number
K760093
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 04/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/24/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number165814
Was Device Available for Evaluation? No
Date Manufacturer Received04/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-