• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS INTUITY ELITE VALVE SYSTEM; TISSUE, HEART-VALVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS INTUITY ELITE VALVE SYSTEM; TISSUE, HEART-VALVE Back to Search Results
Model Number 8300AB
Device Problems Obstruction of Flow (2423); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Complete Heart Block (2627); No Code Available (3191)
Event Date 02/13/2020
Event Type  Injury  
Manufacturer Narrative
Peri-procedural arrhythmias, heart block, and other conduction disturbances are common in patients with underlying cardiovascular disease and/or conduction abnormalities.They can be exacerbated with standard perioperative medications, anesthesia, and/or instrumentation of the heart.Temporary pacemakers are inserted in all patients undergoing aortic valve replacement (avr).It is not uncommon for patients to have short term/reversible periods of heart block or arrhythmias following the procedure while their heart recovers from cardiopulmonary bypass.In many cases, the temporary pm is left in the patient for a short time following the procedure and then subsequently removed prior to discharge.Following surgical aortic valve replacement (avr), new-onset bundle branch block has been reported in 16% to 32% of patients and the need for permanent pacemakers in 3% to 8% of patients.The reason for post-operative av block after surgical avr is related to injury to the cardiac conduction system during surgical excision of the adjacent diseased valve and annular tissue.The close anatomical relationship between the aortic valvular complex and the branching atrioventricular bundle explains the possible development of conduction abnormalities following prosthetic aortic valve procedures.(ref: ts39236) arrhythmias and conduction disturbances are also common after surgical repair or replacement of the tricuspid and mitral valves due to the proximity of the conduction pathways.The device was not returned for evaluation, as it remains implanted.Minimal information regarding this procedure was received and attempts to get additional information regarding the condition of the device, patient's medical history, or possible comorbidities have been unsuccessful.The root cause of the event remains indeterminable.If new information becomes available, a supplemental report will be submitted.The device history record (dhr) was not able to be reviewed as the device serial number was not provided.Edwards will continue to review and monitor all events.Trends are monitored on a monthly basis and if action is required, appropriate investigation will be performed.
 
Event Description
Through review of slides "rapid deployment aortic replacement (radar) registry in (b)(6): initial results.(b)(6)" presented at the heart valve society meeting held in (b)(6), the following events were identified as pertaining to edwards aortic valves: 10 episodes of postoperative atrioventricular block (pavb).
 
Manufacturer Narrative
Reference capa-20-00141.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS INTUITY ELITE VALVE SYSTEM
Type of Device
TISSUE, HEART-VALVE
Manufacturer (Section D)
EDWARDS LIFESCIENCES
one edwards way
irvine CA 92614
MDR Report Key9875952
MDR Text Key194920227
Report Number2015691-2020-11124
Device Sequence Number1
Product Code LWR
Combination Product (y/n)N
PMA/PMN Number
P150036
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 02/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/24/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number8300AB
Was Device Available for Evaluation? No
Date Manufacturer Received07/23/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
-
-