Event Description
On february 28, 2020, nakanishi became aware of a handpiece overheating through a complaint input into the complaint database by a distributor (nsk america).Details are as follows: the event occurred on (b)(6) 2020.A dentist was performing a 3rd molar extraction on the patient using the sga-e2s handpiece (serial no.(b)(4)).The patient was under general anesthesia.During the procedure, the dentist found a blister approximately the size of a pea on the corner of the patient's right lower lip.The injury was cleaned and neosporin was administered to the area.The dentist has conducted a follow up with the patient and the injury has healed normally.No further medical treatment was required for this injury.
 
Manufacturer Narrative
Upon receiving the device involved in the mdr event from the distributor, nakanishi conducted a failure analysis of the returned device that included measuring the operating temperature of the device [report no.(b)(4) ].These activities are described in more detail below.Methodology used: a) nakanishi examined the device history record and the repair history for the subject sga-e2s device [serial no.(b)(6) ].There were no problems observed during manufacturing or testing noted in the dhr.There were also no repair history records since the device was shipped.B) nakanishi conducted temperature testing of the returned device in the following manner: b.1) temperature sensors were attached to the exterior of the device at various test points.This included the point most proximal to the patient (testing point (1)) and points further toward the distal end of the device (testing points (2) through (4)).The test setup was prepared to take temperature measurements at all points simultaneously, including a reference measurement at ambient room temperature.B.2) nakanishi attached a thermocouple (sensor to measure temperature) to each of the testing points.Nakanishi rotated the device's motor at 40,000 min-1, which is the maximum rpm for the motor that drives the handpiece (80,000 min-1 for the handpiece), with water spray, and measured the exothermic response.B.3) nakanishi measured the temperature rise of the returned handpiece set at 80,000 min-1 (motor revolution 40,000 min-1).Nakanishi observed an abnormal temperature rise at test points (1) and (2) a few seconds into the test.Temperature measurements 51 seconds after the start of the test were as follows: - test point (1): 79.4 degrees c - test point (2): 65.5 degrees c - test point (3): 30.4 degrees c - test point (4): 30.3 degrees c the rise in temperature was so sudden that the test was concluded 51 seconds into the planned 5-minute evaluation period.Identification of the specific failure mode(s) and/or mechanism(s) of the associated device components was conducted as follows: a) nakanishi disassembled the handpiece and performed a visual inspection of the internal parts.Nakanishi observed a great amount of debris on the bearing and other internal parts.B) nakanishi took photographs of all of the disassembled parts and kept them in investigation report (b)(4).Conclusions reached based on the investigation and analysis results: a) nakanishi identified that the cause of the overheating of the returned device was abnormal resistance during rotation caused by the soiled bearing.The ingress of undesirable materials caused the bearing to be soiled.B) a lack of maintenance caused the accumulation of debris on the internal parts, which caused debris ingress into the bearing during rotation.This contributed to the handpiece overheating.C) in order to prevent a recurrence of the handpiece overheating, nakanishi took the following actions: c.1) nakanishi reviewed the operation manual and reconfirmed the clarity and understandability of the instructions.C.2) nakanishi reported the above evaluation results to nsk america and directed nsk america to remind the user of the importance of maintenance, as instructed in the operation manual.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NSK
Type of Device
HANDPIECE, ROTARY BONE CUTTING
Manufacturer (Section D)
NAKANISHI INC.
700 shimohinata
kanuma-shi, tochigi-ken 322-8 666
JA  322-8666
MDR Report Key9876451
MDR Text Key196769694
Report Number9611253-2020-00012
Device Sequence Number1
Product Code KMW
Combination Product (y/n)N
PMA/PMN Number
K171155
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Remedial Action Recall
Type of Report Initial,Followup
Report Date 04/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/25/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberSGA-E2S
Device Catalogue NumberH265
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/25/2020
Date Manufacturer Received04/13/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction Number9611253-060818-001-R
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age17 YR
Patient Weight59
-
-