MAUDE Adverse Event Report: STRYKER MEDICAL TRURIZE CHAIR POWERED CLINICAL CHAIR; CHAIR, POSITIONING, ELECTRIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL TRURIZE CHAIR POWERED CLINICAL CHAIR; CHAIR, POSITIONING, ELECTRIC Back to Search Results
Model Number 3333000000
Device Problems Device Alarm System (1012); Material Twisted/Bent (2981)
Patient Problems Fall (1848); Hematoma (1884); Hemorrhage, Subarachnoid (1893)
Event Date 01/06/2020
Event Type  Injury  
Event Description
On the date of the event, icu pt was assisted up in the trurize chair by an rn and pct.The pt's nurse had tested the chair while she was in the pt's room by sitting in the chair and getting up.The chair alarm sounded for the test.The chair was plugged into the call system as per proper procedure.After the nurse tested the chair, she assisted the pt with breakfast and then left the room.Approx 90 mins after the pt was assisted up, a loud thud was heard in the hallway outside of the pt's room by the clinical mgr of icu.The pt had exited the chair, and the call system alarm did not sound in the hallway.The pt was taken to radiology for a ct of the head and neck.The pt was found to have a large hematoma and small subarachnoid hemorrhage.Pt was nonverbal when the incident occurred but became verbal within a few mins.Two subsequent ct scans of the head were completed the 2 days following the incident, that showed decreasing size of the hematoma and stable sah.Upon inspection of the trurize chair after the event, the cord that attached it to the staff call system was found to have a bent prong.The bent prong did not allow the chair exit alarm to be connected to the call system.The chair was taken immediately out of service and the clinical engineering dept fixed the faulty prong.The other mfr was notified of the event along with the other nursing dept that had trurize chairs to let know of the issue with the prong.Fda safety report id# (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TRURIZE CHAIR POWERED CLINICAL CHAIR
Type of Device
CHAIR, POSITIONING, ELECTRIC
Manufacturer (Section D)
STRYKER MEDICAL
MDR Report Key9877736
MDR Text Key185084071
Report NumberMW5093914
Device Sequence Number1
Product Code INO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Risk Manager
Type of Report Initial
Report Date 03/23/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/24/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3333000000
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age72 YR
Patient Weight60
-
-