The investigation has determined that higher than expected lactate results were obtained from a vitros performance verifier (pv) quality control (qc) fluid using vitros chemistry products lac slides lot 3532-0108-3860 on a vitros 350 chemistry system.A definitive assignable cause could not be determined.Based on historical quality control results, a vitros lac reagent lot 3532-0108-3860 performance issue is not a likely contributor to the event.Furthermore, continual tracking and trending of complaint data has not identified any signals to suggest there is a systemic quality issue with vitros lac reagent lot 3532-0108-3860.However, the customer exhausted their supply of vitros lac lot 3532-0108-3860 and no additional troubleshooting using this lot was able to be performed.The issue was resolved when using an alternate lot of vitros lac reagent.Therefore, an issue related to vitros lac lot 3532-0108-3860 could not be completely ruled out nor confirmed as a contributing factor of the event.Although precision testing was not performed on the vitros 350 chemistry system an instrument related issue is not a likely contributor of the event as historical quality control results were precise and an alternate lot of vitros lac reagent yielded acceptable results in combination with the vitros 350 chemistry system without any troubleshooting actions being performed on the instrument.Because fresh vials of the vitros pv i and pv ii control fluids were used as part of troubleshooting by the customer, an issue related to the vitros pv fluids cannot be ruled out nor confirmed as a contributor of the event.
|