Catalog Number UNK HIP FEMORAL HEAD |
Device Problem
Corroded (1131)
|
Patient Problems
Foreign Body Reaction (1868); No Code Available (3191)
|
Event Date 04/09/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
|
|
Event Description
|
It was reported that the patient had metal on metal revision.New liner and head implanted.Doi: unk.Dor: (b)(6) 2019; left hip.
|
|
Manufacturer Narrative
|
Product complaint # (b)(4).Investigation summary: the device associated with this report was not returned to depuy synthes for evaluation.All available x-rays were reviewed, and no evidence of implant fracture, disassociation, or anything indicative of a device nonconformance was found, not confirming the reported allegation.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.Device history lot: a device history record (dhr) review or manufacturing records evaluation (mre) was not possible because the investigation was not able to retrieve the required lot code.
|
|