• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION CAPTIVATOR; SNARE, FLEXIBLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION CAPTIVATOR; SNARE, FLEXIBLE Back to Search Results
Model Number M00561311
Device Problems Use of Device Problem (1670); Detachment of Device or Device Component (2907)
Patient Problem Burn(s) (1757)
Event Date 01/28/2020
Event Type  Injury  
Manufacturer Narrative
(b)(4).According to the complainant, the suspect device has been disposed and is not available for return.If any further relevant information is received, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that a captivator large oval med stiff snare was used in the colon during a polypectomy procedure performed on (b)(6) 2020.According to the complainant, during procedure and outside the patient, the snare was passed several times in and out of the scope and at some point, the catheter got detached from the handle leaving the wire at the top exposed.When the snare was removed, the exposed wire was not noticed and was placed externally on the patient.While the endoscopist was removing the scope, he stepped on the diathermy pedal and the snare burned the patient's leg.The procedure was completed with this device.Boston scientific has been unable to obtain additional information regarding the event to date, despite good faith efforts.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CAPTIVATOR
Type of Device
SNARE, FLEXIBLE
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC DE COSTA RICA S.R.L.
2546 calle primera
propark, coyol
alajuela
CS  
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key9903894
MDR Text Key189578648
Report Number3005099803-2020-01412
Device Sequence Number1
Product Code FDI
UDI-Device Identifier08714729019251
UDI-Public08714729019251
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K131700
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 03/31/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received03/31/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date06/24/2021
Device Model NumberM00561311
Device Catalogue Number6131
Device Lot Number0022297845
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/09/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured06/25/2018
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-