MEDTRONIC EUROPE SARL CLARIA MRI QUAD CRT-D SURESCAN; DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
|
Back to Search Results |
|
Model Number DTMA2QQ |
Device Problems
Loose or Intermittent Connection (1371); Mechanical Problem (1384); Decreased Sensitivity (2534)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 03/13/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported that during implant the p-wave measured normal through the analyzer, however, once the right atrial (ra) lead was connected to the device, the p-waves were low.The physician suspected it did not connect well so they reconnected the lead but the p wave did not improve.The physician measured the p-wave through the analyzer again and it was normal.When the lead was connected back to the device, it was low again.The lead was repositioned with normal p waves, however, when the physician tried to reconnect back to the device, the set screw was crooked and they were unable to connect the lead to the device.A new device was used.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the device was returned and analyzed.Returned product analysis was performed and no anomalies were found.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|
|
|