(b)(4).Batch #t94u1h.Investigation summary: the device was returned with the distal tip of the blade broken off and returned with the device.The remaining blade portion was scratched with evidence of contact with metal in or out of the operative field.This blade tip portion may have broken off of the device during transport to our analysis site.During functional testing on gen11, an alert screen was displayed.A probable cause for the device to stop activating and the gen11 to display an alert screen is blade damage.The device was disassembled to inspect the internal components and no anomalies were found.The manufacturing records were reviewed and the manufacturing/packaging criteria were met prior to the release of this batch.Probable causes of blade damage, including breakage, are external contact during pre-op or general use, blade contact with other devices, staples or clips during the procedure.Once minor blade damage has occurred, subsequent activations may increase the severity of the blade damage.This in turn can result in activation issues such as failing the pre-run test with the generator and displaying an alert screen.These alert screens that can result include "remove instrument from patient," ¿blade error detected¿ or "relax pressure on blade," followed by a ¿replace instrument¿ screen later in the procedure.Continued usage of the damage blade can result in a broken blade.
|
It was reported that during a laparoscopic hepatectomy, "activate instrument outside patient to run test" was displayed when the device was used on the liver surface.A pre-run test was done again, but the device did not pass the test.No metal clip was used with the device.The blade tip was not broken off and no pieces fell into the patient.Gen11 was used.Another device was used to complete the case.There were no adverse consequences to the patient.No further information is available.
|