The patient's sample was sent to the manufacturer for investigation.No interfering factor could be detected in the sample.The measured autoantibodies in a patient¿s sample can vary depending on the testing procedure used.Autoantibody values measured in patient samples by different testing procedures cannot be directly compared.Elecsys® anti-ccp is a second generation anti-ccp assay.More than one cyclic citrullinated peptide is used.In this assay, four different artificial cyclic citrullinated peptides are used.The amount of given cyclic citrullinated peptides in the elecsys® assay may lead to a higher sensitivity.Reactivity of autoantibodies in different tests vary, because of the use of different assay formats, assay reagents, detection antibodies, antigen formulations and incubation times.Product labeling states, "anti-ccp values determined on patient samples by different testing procedures cannot be directly compared with one another and could be the cause of erroneous medical interpretations." the investigation did not identify a product problem.The cause of the event could not be determined.
|