DAVIS & GECK CARIBE LTD ENDO CLIP; CLIP, IMPLANTABLE Back to Search Results
Model Number 176625
Device Problems Mechanics Altered (2984); Noise, Audible (3273)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/24/2020
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
According to the reporter, during partial gastrectomy, at the third firing, cracking sound was heard and the clip failed to come out.A new product was used and it was able to be used without problem.There was no patient injury.
 
Manufacturer Narrative
Evaluation summary: post market vigilance (pmv) led an evaluation of one device.The instrument was received partially applied with twelve remaining clips.During functional evaluation, it was observed that the jaws would not close upon actuation of the handle.The instrument body was removed from the shaft and disassembled for visualization of internal components.It was observed that the firing handle linkage was forced through the channel tube lugs, suggesting that the instrument had been fired through the safety interlock.Due to the noted damage, further functional testing of the instrument could not be performed.In addition; the interlock mechanism was triggered as a result of rapidly actuating the handles of the device and or forcing the device, this renders the device inoperable and poses no harm to the patient.Records from each manufacturing lot are thoroughly reviewed to ensure that products are released meeting all quality release specifications at the time of manufacture.Replication of this condition may occur when an attempt is made to fire the instrument without first loading a clip by using the trigger.The safety interlock feature is designed to prevent the jaws from closing on a vessel in the absence of a loaded clip.If an attempt is made to forcibly fire the instrument while engaged in interlock, the lugs are designed to break as noted and the interlock feature continues to function as intended.No further actions have been deemed necessary at this time.The root cause of the observed damage was found to be due to the device not being used as intended which caused or contributed to the reported condition.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENDO CLIP
Type of Device
CLIP, IMPLANTABLE
Manufacturer (Section D)
DAVIS & GECK CARIBE LTD
zona franca de san isidro
santo domingo 18750
DO  18750
MDR Report Key9937257
MDR Text Key187456030
Report Number9612501-2020-00624
Device Sequence Number1
Product Code FZP
UDI-Device Identifier20884521057842
UDI-Public20884521057842
Combination Product (y/n)N
PMA/PMN Number
K954435
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 05/15/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/08/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number176625
Device Catalogue Number176625
Device Lot NumberJ9H1672NY
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/08/2020
Date Manufacturer Received04/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-