VYAIRE MEDICAL 3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV); VENTILATOR, HIGH FREQUENCY Back to Search Results
Model Number 3100 A
Device Problem Overheating of Device (1437)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 03/12/2020
Event Type  Injury  
Manufacturer Narrative
Vyaire field service representative (fsr) went onsite and evaluated the ventilator.Fsr noticed excessive amount of dust on the filter located at the bottom rear of the oscillator stand.No other failures found.Unit most likely overheated and shutdown.Fsr cleaned the filter, verified that the alarm board is ok and the thermistor was not damaged.Fsr checked all settings on the power width modulator.Verified that the ddi board is good.Checked all voltages coming from the power supply.Unit passed all calibrations and can be returned to service.Vyaire medical will submit a supplemental report in accordance with 21 cfr section 803.56 if additional information was received.
 
Event Description
It was reported to vyaire medical that the 3100a unit shuts off and would not power back on during patient use.The patient was transferred to another ventilator.The customer reported that there was no patient harm during this incident.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
3100 HIGH FREQUENCY OSCILLATING VENTILATOR (HFOV)
Type of Device
VENTILATOR, HIGH FREQUENCY
Manufacturer (Section D)
VYAIRE MEDICAL
26125 n. riverwoods blvd.
mettawa, il
Manufacturer (Section G)
VYAIRE MEDICAL INC.
1100 bird center drive
palm springs, ca
Manufacturer Contact
mindy faber
26125 n. riverwoods blvd.
mettawa, il 
MDR Report Key9945289
MDR Text Key187212694
Report Number2021710-2020-11672
Device Sequence Number1
Product Code LSZ
UDI-Device Identifier10846446003079
UDI-Public(01)10846446003079(11)20090201
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P890057
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 03/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/09/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number3100 A
Device Catalogue Number768901
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Date Manufacturer Received03/12/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2009
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-