One single dpt kit with iv set and pressure tubing arrived.Priming solution was visible throughout the kit.A non-edwards (terumo) syringe was attached at zero-stopcock.The reported event of contamination - white foreign material, was not confirmed.There were no visible particulates or contamination that were observed from the returned kit, including non-ew syringe which was attached at zero-stopcock.The plunger of non-edwards syringe was pushed to closed position and then the fluid path of the returned kit was flushed continuously for 5 minutes, however no visible particulates were flushed out from the kit.Additional evaluation task has assigned to manufacturing site for further investigation. a device history record review was completed and documented that device met all specifications upon distribution.It is common clinical practice to inspect all products before usage.Additionally, these products are used by highly trained clinicians, experienced in identifying and mitigating any hazards that arise.Invasive procedures involve some patient risks.Although serious complications are relatively uncommon, the physician is advised to consider the potential benefits in relation to the possible complications.The noted particulate was not confirmed during visual inspection.However, complaint histories for all reported events are reviewed against trending control limits on a monthly basis and any excursions above the control limits are assessed and documented as a part of the monthly review.
|