• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MOOG MEDICAL DEVICES GROUP ENTERALITE INFINITY FEEDING PUMP; ENTERAL INFUSION PUMP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MOOG MEDICAL DEVICES GROUP ENTERALITE INFINITY FEEDING PUMP; ENTERAL INFUSION PUMP Back to Search Results
Model Number INFKIT2
Device Problem Free or Unrestricted Flow (2945)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/20/2020
Event Type  malfunction  
Manufacturer Narrative
The device was not returned to mmdg for investigation.A dhr was completed and found no non-conformances.Because the device was not returned to mmdg the complaint could not be investigated or confirmed.This report will be updated if the device is returned to mmdg.
 
Event Description
The initial reporter stated that the pump was alarming no flow out.They stated that they massaged the tubing and in-line occluder and that the pump started to free flow.They stated that they removed the administration set, and placed it back in the pump and that it stopped free flowing.Mmdg did follow up with the initial reporter who stated that the patient was doing well and there had been no adverse events associated with the complaint.(b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENTERALITE INFINITY FEEDING PUMP
Type of Device
ENTERAL INFUSION PUMP
Manufacturer (Section D)
MOOG MEDICAL DEVICES GROUP
4314 zevex park lane
salt lake city, ut
Manufacturer (Section G)
MOOG MEDICAL DEVICES GROUP
4314 zevex park lane
salt lake city, ut
Manufacturer Contact
kristin hardesty
4314 zevex park lane
salt lake city, ut 
2641001112
MDR Report Key9957579
MDR Text Key187556363
Report Number1722139-2020-00127
Device Sequence Number1
Product Code LZH
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K031199
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Nurse Practitioner
Type of Report Initial
Report Date 03/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberINFKIT2
Device Catalogue NumberINFKIT2
Device Lot NumberN/A
Was Device Available for Evaluation? No
Date Manufacturer Received03/20/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/26/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age3 YR
-
-