• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUICKSITE LV; PERMANENT PACEMAKER ELECTRODE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUICKSITE LV; PERMANENT PACEMAKER ELECTRODE Back to Search Results
Model Number 1056T/86
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem No Information (3190)
Event Date 04/01/2020
Event Type  malfunction  
Manufacturer Narrative
No complaint was received with return of the device.Failure event observed during analysis.Final analysis found only the distal portion was returned in one piece measuring 70.4 cm.Internal insulation abrasion breaching the ring electrode (re) lumen was noted at 5.0 ¿ 5.5 cm from the distal tip.The etfe coating of the re cable was intact in this region.Other damages found were consistent with surgical damage.
 
Event Description
This report is to advise of an event observed during analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QUICKSITE LV
Type of Device
PERMANENT PACEMAKER ELECTRODE
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
elizabeth boltz
15900 valley view court
sylmar, CA 91342
MDR Report Key9959722
MDR Text Key187617533
Report Number2017865-2020-05023
Device Sequence Number1
Product Code OJX
Combination Product (y/n)N
PMA/PMN Number
P030054
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other
Type of Report Initial
Report Date 04/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/14/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date12/31/2008
Device Model Number1056T/86
Device Lot Number0002118005
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer03/04/2020
Was the Report Sent to FDA? No
Date Manufacturer Received04/01/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/17/2005
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-