• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065751462
Device Problem Device Slipped (1584)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/30/2020
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A customer reported that a piece of the trocar cannula fell into the eye and removed during a procedure.The procedure was completed with no patient impact.
 
Manufacturer Narrative
Additional information provided in d.10., h.3., h.6., and h.10.One opened total plus pak with various components was received but no trocar assemblies were returned for the reported issue of the parts of the product fell to the eyeball; therefore, the condition of the product could not be verified.A review of the device history record traceable to the reported lot number, indicates that the product was processed and released according to the product¿s acceptance criteria.A sample was not received at the manufacturing site and the device history record traceable to the reported lot number indicates that the product was processed and released according to the product¿s acceptance criteria, therefore, the root cause for the customer complaint issue cannot be determined.The exact root cause for this complaint is unknown, therefore, specific action with regards to this complaint cannot be taken.An acceptance quality inspection is performed to ensure product meets release acceptance criteria.Non-conforming product is removed from the lot.The inspection includes evaluation for trocar hub/cannula assembly loose on the trocar blade.Complaints are reviewed and monitored at regular intervals for any significant adverse trends.No additional action is required at this time the manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
MDR Report Key9968437
MDR Text Key188218369
Report Number1644019-2020-00208
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 07/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/31/2021
Device Catalogue Number8065751462
Device Lot Number2305476H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/04/2020
Date Manufacturer Received06/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-