• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS HEALTHCARE DIAGNOSTICS INC. CLINITEK STATUS+

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS HEALTHCARE DIAGNOSTICS INC. CLINITEK STATUS+ Back to Search Results
Model Number 10379675
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Patient Involvement (2645)
Event Date 03/25/2020
Event Type  malfunction  
Manufacturer Narrative
Siemens has requested the instrument and batteries to be returned for investigation.The customer has received a replacement instrument under warranty and they are operational.The cause of this event is unknown.
 
Event Description
The customer reported that when they inserted a new set of batteries into the instrument, it started smoking.There was no report of injury to user or property due to the event.
 
Manufacturer Narrative
Siemens evaluated the clinitek status+ returned by the customer.The instrument was returned in good condition with no damage to the exterior of the instrument.The instrument was returned without batteries or the brick portion of the power supply (cable only).The instrument has d16 damage on the main board.D16 failure caused the system to not operate.The d16 was the source of the burning smell but it is not a fire risk.D16 fails when incorrect power is connected to the device, either through plugging in the wrong power supply (device uses a common barrel connector) or using 3.5v rechargeable batteries (device specifies using common 1.5v aa batteries).The customer is operational with the new instrument.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CLINITEK STATUS+
Type of Device
CLINITEK STATUS+
Manufacturer (Section D)
SIEMENS HEALTHCARE DIAGNOSTICS INC.
511 benedict avenue
tarrytown, ny
MDR Report Key9969687
MDR Text Key188658930
Report Number3002637618-2020-00018
Device Sequence Number1
Product Code JIL
UDI-Device Identifier00630414574639
UDI-Public00630414574639
Combination Product (y/n)N
PMA/PMN Number
K091216
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Remedial Action Replace
Type of Report Initial,Followup
Report Date 06/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/16/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10379675
Device Catalogue Number10379675
Was Device Available for Evaluation? Yes
Date Manufacturer Received05/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age35 YR
-
-