Catalog Number 00515047501 |
Device Problems
Detachment of Device or Device Component (2907); Material Split, Cut or Torn (4008)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 03/19/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).Once an investigation of the device is completed, a follow-up/final report will be submitted.Initial reporter- telephone number: (b)(6).
|
|
Event Description
|
It was reported that during the surgery, the radial tip came off from the handpiece easily during use.Additionally, the cord was cut on the handpiece.There was no harm, but there was a 0-15 minute delay to prepare an alternate product.No adverse events were reported as a result of this malfunction.
|
|
Manufacturer Narrative
|
This event has been recorded by zimmer biomet under (b)(4).This medwatch is being filed to relay additional information.Visual examination of the provided pictures showed that the cord/tubing was cut, but could not identify the tip coming off.Review of complaint history identified additional similar complaints for the reported item and part and lot combination.Complaints are monitored through monthly complaint review (reference (b)(4)) in order to identify potential adverse trends.Device is used for treatment.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is not confirmed.H3 other text : product location unknown.
|
|
Event Description
|
There is no additional information.
|
|