• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ADVANCED BIONICS, LLC HIRES ULTRA IMPLANT; COCHLEAR IMPLANT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ADVANCED BIONICS, LLC HIRES ULTRA IMPLANT; COCHLEAR IMPLANT Back to Search Results
Model Number CI-1600-05
Device Problems Appropriate Term/Code Not Available (3191); Intermittent Loss of Power (4016)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 03/27/2020
Event Type  malfunction  
Event Description
The patient was reportedly experiencing intermittencies.The patient's device was explanted.The patient was reimplanted with another advanced bionics cochlear device.
 
Manufacturer Narrative
Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The recipient was reimplanted with another cochlear device.The external visual inspection revealed silicone slices on the top cover, near the fantail and along the lead prior to receipt.These are believed to have occurred during revision surgery.The device passed the photographic imaging inspection.System lock was verified.The electrode condition prevented an electrical test from being performed.The device passed the electrical tests performed.Advanced electrical testing performed on the device revealed lower than typical range on some of the electrodes.It is the opinion of advanced bionics explant review board that some of these electrode test results should be considered as failures.The device passed the mechanical test performed.The failure of this device is attributed to an electrode short in the electrode pocket.A corrective action was implemented.This version of the ultra device is no longer distributed.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES ULTRA IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key9972939
MDR Text Key188076768
Report Number3006556115-2020-00277
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016844584
UDI-Public(01)07630016844584(11)171205(17)201130
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup,Followup
Report Date 04/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/17/2020
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/30/2020
Device Model NumberCI-1600-05
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age63 YR
-
-