This event has been recorded by zimmer biomet under: (b)(4).This medwatch is being filed to relay additional information.Product review of the air dermatome serial number: (b)(6) by a zimmer biomet certified service repair technician on 24 april 2020 revealed that the motor was erratic, the lever was loose, the width plate was damaged and the device was out of calibration.Repair of the device was performed by a zimmer biomet certified service repair technician on 24 april 2020 which included replacement of the lever, and width plate.The device, serial number: (b)(6), was then tested and functioned properly.It was repaired, inspected and tested.Device is used for treatment.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.The event is confirmed.
|