Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem
No Patient Involvement (2645)
Event Type
malfunction
Event Description
Information was received indicating that cadd solis hpca pump's volume measure was out of calibration.There was no patient involved.
Manufacturer Narrative
One cadd solis hpca pump was returned for analysis in good condition.No evidence of fault exists upon review of the event history log.Accuracy testing was performed; fault not duplicated.Based on the evidence, there was no fault found as the complaint was not confirmed.