|
Model Number ENO DR |
Device Problems
Difficult to Interrogate (1331); Pacemaker Found in Back-Up Mode (1440)
|
Patient Problem
No Information (3190)
|
Event Date 03/31/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The device involved in this mdr report is not approved in the united states; however, it is similar to a device manufactured by sorin that was cleared or approved by fda for marketing in the united states.
|
|
Event Description
|
Reportedly, the patient had intermittent complete av block.During implantation, when the leads were connected to the pacemaker, normal pacing was observed at 60 min-1.However, a few minutes later, the device switched in vvi mode at 70 min-1 and unipolar pacing until the end of the procedure.The physician used electrocautery to close the pocket.After the procedure, interrogation of the pacemaker was not possible.The programmer continued to load for a while, with the word 'interrogation' at the top of the screen.As interrogation was not possible, and no buttons on the programmer could be pressed, the programmer was restarted.Upon re-interrogation of the pacemaker, a message stating that the device was in standby mode was displayed.After the user pressed 'ok', interrogation was still not possible.The red cross button on the programmer keyboard was then pressed and the pacemaker could be interrogated.
|
|
Event Description
|
Reportedly, the patient had intermittent complete av block.During implantation, when the leads were connected to the pacemaker, normal pacing was observed at 60 min-1.However, a few minutes later, the device switched in vvi mode at 70 min-1 and unipolar pacing until the end of the procedure.The physician used electrocautery to close the pocket.After the procedure, interrogation of the pacemaker was not possible.The programmer continued to load for a while, with the word 'interrogation' at the top of the screen.As interrogation was not possible, and no buttons on the programmer could be pressed, the programmer was restarted.Upon re-interrogation of the pacemaker, a message stating that the device was in standby mode was displayed.After the user pressed 'ok', interrogation was still not possible.The red cross button on the programmer keyboard was then pressed and the pacemaker could be interrogated.
|
|
Manufacturer Narrative
|
Please refer to the attached analysis report.
|
|
Search Alerts/Recalls
|
|
|