• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BLUE BELT TECHNOLOGIES NAVIO TKA PLANE VISUALIZATION TOOL; ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BLUE BELT TECHNOLOGIES NAVIO TKA PLANE VISUALIZATION TOOL; ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number ROB00004
Device Problem Fitting Problem (2183)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/21/2019
Event Type  malfunction  
Event Description
It was reported that plane visualization tool for needs to be replaced because they are all snug or get stuck when affixing them to the handpiece with the drill inserted.Drills need to be removed in order to properly attach.No patient involved.
 
Manufacturer Narrative
The devices were being used during treatment and were returned for evaluation.The plate probes fit over all of the handpieces without issue.The serial number for the device was not provided.However, all lots of pn (b)(4).Distributed to the field from when the part number was first inspected ((b)(6)2016) to the complaint occurrence date ((b)(6)2019) were reviewed finding one lot that was related to fit issues between the handpiece and the visualization tool.A complaint history review was performed and found reports of the issue.This issue will continue to be monitored.A relationship between the device and the reported event could be established.Functional evaluation of the plate probes found that they fit over the handpieces without any issues.A factor that could have contributed to the issue is if there was debris caught between the handpiece and visualization tool, preventing it from being fully locked in.However, there was no problem found.No containment or corrective actions are recommended at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
NAVIO TKA PLANE VISUALIZATION TOOL
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
BLUE BELT TECHNOLOGIES
2905 northwest blvd ste 40
plymouth MN 55441
MDR Report Key9999433
MDR Text Key189165456
Report Number3010266064-2020-01433
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00885556685068
UDI-Public00885556685068
Combination Product (y/n)N
PMA/PMN Number
K191223
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/13/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received04/26/2020
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberROB00004
Device Catalogue NumberROB00004
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/14/2019
Date Manufacturer Received08/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-