In response to cardioquip's "urgent medical device correction mch-1000 cooler heater" it was determined that our 3 cardioquip mch-1000 cooler heaters are contaminated.This was verified through visual inspection and water testing.On (b)(6) 2022 heterotophic plate count, sm 9215b - 28000.Test unit: unk.Low test range: na.High test range: 500 cfu/ml.Our cleaning practices exceeded the manufacturer recommendation by performing monthly cleaning as opposed to recommended quarterly cleaning.We are in the process of returning our mch-1000s to cardioquip to be reconditioned.Fda safety report id # (b)(4).
|