• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH VIPER UNI SCREW 5X35MM TI; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH VIPER UNI SCREW 5X35MM TI; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 186720535
Device Problems Break (1069); Entrapment of Device (1212)
Patient Problems Foreign Body In Patient (2687); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/24/2024
Event Type  malfunction  
Event Description
Device report from synthes reports an event in switzerland as follows: it was reported that a young patient underwent a t12-l2 fixation in (b)(6) 2022 following a burst fracture of l1.Shortly before the normal ablation surgery, about a year later, one found that one screw tip was broken.The screw ablation was performed on (b)(6) 2024, but one screw tip (fragment) was left in the body of t12 vertebra.There was no patient consequence reported.This report is for one (1) viper uni screw 5x35mm ti this is report 1 of 1 for complaint (b)(4).
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.E3: reporter is a j&j employee.D9: complainant part is not expected to be returned for manufacturer review/investigation.H3, h4, h6: without a lot number the device history records review could not be completed.Product was not returned.Based on the information available, it has been determined that no corrective and/or preventative action is proposed.This complaint will be accounted for and monitored via post market surveillance activities.If additional information is made available, the investigation will be updated as applicable.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.H10 additional narrative: h3, h6: a product investigation was conducted.The product was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned device.Visual analysis of the returned sample revealed that the tip of the viper uni screw 5x35mm ti was broken.Photos were provided and embedded device condition can be confirmed due to tip screw fragment can be observed in the patient body.However; with the evidence provided a potential cause cannot be determined.A dimensional inspection for the viper uni screw 5x35mm ti was not performed due to post manufacturing damage.The observed condition of the device was consistent with a random component failure that may have been caused by exposure to unintended forces.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed as the observed condition of the viper uni screw 5x35mm ti would contribute to the complained device issue.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.H3, h4, h6: device history record (dhr) review conducted: part # 186727535.Lot # 293252.Supplier: (b)(4).Batch1: lot qty of (b)(4).Unit were released on 07 apr 2021 with no discrepancies.Batch2: lot qty of (b)(4).Unit were released on 07 apr 2021 with no discrepancies.Batch3: lot qty of (b)(4).Unit were released on 29 apr 2021 with no discrepancies.Batch4: lot qty of (b)(4).Unit were released on 12 apr 2021 with no discrepancies.Batch5: lot qty of (b)(4).Unit were released on 16 apr 2021 with no discrepancies.Batch6: lot qty of (b)(4).Unit were released on 20 apr 2021 with no discrepancies.Batch7: lot qty of (b)(4).Unit were released on 09 apr 2021 with no discrepancies.No ncrs were generated during production.The ri identified no issues during the manufacturing and release of this device that could have contributed to the problem reported by the customer.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.If the information is unknown, not available or does not apply, the section/field of the form is left blank.Device was received, the investigation is in progress, no conclusion could be drawn at the time of filing this report.The photo was returned to depuy synthes for evaluation.The depuy synthes team conducted a visual inspection of the returned photo.Visual analysis of the photo and x-ray image revealed that the tip viper uni screw 5x35mm ti had broken.Embedded device condition can be confirmed due to tip screw fragment can be observed in the patient body.No other issues were identified.With the evidence provided a potential cause cannot be determined.As the device was not returned, an as-received condition could not be assessed, and a dimensional inspection and document/specification review were not completed.As part of depuy synthes quality process, all devices are manufactured, inspected, and released to approved specifications.The overall complaint was confirmed for viper uni screw 5x35mm ti.There is no indication that a design or manufacturing issue has caused the complaint condition.Based on the investigation findings, it has been determined that no corrective and/or preventative action is proposed.Additional monitoring for any potential safety signals will be conducted through complaint trending and other post-market safety surveillance activities.The device lot number is unknown, therefore a device history review could not be performed.If the lot/serial number becomes available, the record will be re-assessed.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VIPER UNI SCREW 5X35MM TI
Type of Device
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
SZ  
Manufacturer Contact
kate karberg
chemin-blanc 38
le locle 
SZ  
3035526892
MDR Report Key18650012
MDR Text Key334884069
Report Number1526439-2024-00823
Device Sequence Number1
Product Code NKB
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K160879
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 02/06/2024
2 Devices were Involved in the Event: 1   2  
1 Patient was Involved in the Event
Date FDA Received02/06/2024
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number186720535
Device Lot Number293252
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? Yes
Date Manufacturer Received04/16/2024
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured04/07/2021
Is the Device Single Use? Yes
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age14 YR
Patient SexFemale
-
-