There was an allegation of a questionable gluc hk gen.3 (glucose) result from the cobas 8000 c702 module.The initial result of 1.67 mmol/l, accompanied by a data flag, and automatically repeated as 2.71 mmol/l.The repeat analyses performed from the same tube (not recentrifuged) on two other laboratory instruments confirmed the lower initial result.The exact result was not provided and was reported outside of the laboratory.
|
On (b)(6) 2024, the controls were within the expected range, and on (b)(6) 2024, the controls were higher than the expected range.An allegation of new questionable results (patient 02 and patient 03) was reported by the customer.The new measurements were performed with reagent gluc hk gen.3 lot no.772291 on (b)(6) 2024: patient 02: the initial glucose result was 1.08 mmol/l, accompanied by a data flag, and repeated as 5.91 mmol/l and 5.74 mmol/l.Patient 03: the initial glucose result was 28.19 mmol/l, and repeated as 8.51 mmol/l, 27.71 mmol/l, and 28.31 mmol/l.The investigation is ongoing.
|