The initial reporter stated they received discrepant results for two patient samples tested with the na electrode on a cobas 6000 c 501 analyzer.No questionable results were reported outside of the laboratory.The samples were repeated due to the values being considered critical.The first sample initially resulted in a na value of 232 mmol/l and when repeated on a different analyzer, the result was 140 mmol/l.The second sample initially resulted in a na value of 252 mmol/l and it repeated as 202 mmol/l.When repeated on a different analyzer, the result was 142 mmol/l.The 142 mmol/l value was deemed correct.
|
The field service engineer found a hole in the vacuum tubing, causing cells to overflow.The tubing was replaced and the tubing bundle was adjusted so that it does not rub.The vacuum pump diaphragms were checked, valves were flushed, and mechanism checks were performed.The reagent probe was adjusted and the system was purged of air.A hardware check test and ise checks passed.The customer ran calibration and controls with passing results.Precision studies were performed.The investigation determined the service actions resolved the issue.Medwatch fields d1, d2, d4, g1, and g4 have been updated.
|